To study the effects of 5.6 GHz radiofrequency irradiation on heart rate, blood pressure, and respiratory rate in anesthetized rats during intermittent exposures which resulted in 1°C colonic temperature changes.
Exposure | Parameters |
---|---|
Exposure 1:
5.6 GHz
Modulation type:
CW
Exposure duration:
intermittent*
|
|
Exposure 2:
5.6 GHz
Modulation type:
pulsed
Exposure duration:
intermittent*
|
|
*The exposure was preformed until the colonic temperature of 39.5°C was reached, then stoped until it is 38.5°C and iniated again until it reached 39.5°C. This procedure was repeated for eight cycles
Frequency | 5.6 GHz |
---|---|
Type | |
Waveform | |
Charakteristic | |
Exposure duration | intermittent* |
Modulation type | CW |
---|
Measurand | Value | Type | Method | Mass | Remarks |
---|---|---|---|---|---|
power density | 600 W/m² | mean | measured | - | 12 W/kg corresponding SAR value; exposure duration of 6.3 ± 0.3 min |
power density | 300 W/m² | mean | measured | - | 6 W/kg corresponding SAR value; exposure duration of 14.9 ± 1.2 min |
Frequency | 5.6 GHz |
---|---|
Type | |
Waveform | |
Charakteristic | |
Exposure duration | intermittent* |
Modulation type | pulsed |
---|---|
Additional info |
pulse duration 1 or 2 µs; repetition rates 250 or 500 pps |
Exposure source |
|
---|
Measurand | Value | Type | Method | Mass | Remarks |
---|---|---|---|---|---|
power density | 300 W/m² | mean | measured | - | 6 W/kg corresponding SAR value; exposure duration of 14.9 ± 1.2 min |
power density | 600 W/m² | mean | measured | - | 12 W/kg corresponding SAR value; exposure duration of 6.3 ± 0.3 min |
There was no differences between results of continuous wave and pulsed wave exposure. During irradiation to 60 mW/cm² which resulted in a 1°C change in colonic temperature, heart rate increased; the values returned to control levels after irradiation was discontinued. No changes in mean arterial blood pressure or in respiratory rate were revealed. Exposure to 30 mW/cm² produced no significant changes in heart rate, blood pressure, or respiratory rate.
The results indicate that heart rate changes during exposure to 5.6-GHz radiofrequency irradiation are related to the average power density applied, and thus to the rate of change in temperature.
This website uses cookies to provide you the best browsing experience. By continuing to use this website you accept our use of cookies.